| Customization: | Available |
|---|---|
| Type: | Material Testing Machine |
| Load Way: | Electronic Load |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency


Standards | GB/T 16584, GB/T 33095.1, ASTM D5289, ISO 6502, DIN 53529 |
Frequency | Fixed at 1.67 Hz (100 cpm) or selectable from 0.001 Hz to 50 Hz |
Amplitude (Strain) | Fixed angle ±0.5° (±7%), ±0.2° (±2.80%), ±1.0° (±14%), ±3.0° (±42%) or selectable from 0.001° to 90° (±0.07%~±1256%) |
Temperature | 40~210℃; higher temperature up to 230℃ optional |
Torque Range | Standard 0.01~225 dNm, 0.01~50 dNm optional |
Pressure Range | 8300 kPa (1200 psi) |
Sample Volume | 4.5 cm³ |
Data Points | Typical data ML, MH, Tanδ@ML, Tanδ@MH, ts1, ts2, t10, t50, t90 (any vulcanization time selectable), curing rate, MH-ML, reversion test TR. Pressure point data PL, PH, P10, P50, P90 (optional) |
Automated Sample Loading System | Options for 15, 30, or 120 samples |
Reproducibility & Repeatability of MDR Test Data | |||||||
Lanxess Standard Rubber Test Angle: ±0.5° Test Temperature: 177°C Test Duration: 12 minutes 5 Parallel Laboratory Experiments Results | |||||||
Test Project | S'@ML | S'@MH | TC10 | TC50 | TC90 | S''@ML | S''@MH |
Average Value | 1.4 | 11.59 | 1.41 | 3.07 | 5.31 | 0.91 | 1.02 |
Repeatability Standard Deviation | 0.028 | 0.093 | 0.026 | 0.028 | 0.047 | 0.03 | 0.028 |
Repeatability Coefficient of Variation | 2.00% | 0.80% | 1.84% | 0.91% | 0.89% | 3.29% | 2.75% |
Reproducibility Standard Deviation | 0.069 | 0.13 | 0.078 | 0.078 | 0.173 | 0.088 | 0.086 |
Reproducibility Coefficient of Variation | 4.93% | 1.12% | 5.53% | 2.90% | 3.26% | 9.67% | 8.43% |